一、用途
ZJC-50KV絕緣材料介電擊穿強(qiáng)度測(cè)試儀采用計(jì)算機(jī)控制,通過(guò)人機(jī)對(duì)話方式,完成對(duì)絕緣介質(zhì)的工頻電壓擊穿,工頻耐壓試驗(yàn)。主要適用于固體絕緣材料如絕緣漆、樹(shù)脂和膠、浸漬纖維制品、云母及其制品、塑料、薄膜復(fù)合制品、陶瓷和玻璃等介質(zhì)在工頻電壓或直流電壓下?lián)舸?qiáng)度和耐電壓時(shí)間的測(cè)試;采用計(jì)算機(jī)控制,可對(duì)試驗(yàn)過(guò)程中的各種數(shù)據(jù)進(jìn)行快速、準(zhǔn)確的采集、處理,并可存取、顯示、打印。
二、依據(jù)標(biāo)準(zhǔn)
1、GB1408.1-2006《絕緣材料電氣強(qiáng)度試驗(yàn)方法》
2、GB1408.2-2006《絕緣材料電氣強(qiáng)度試驗(yàn)方法
第2部分:對(duì)應(yīng)用直流電壓試驗(yàn)的附加要求》
3、GB/T1695-2005《硫化橡膠工頻擊穿電壓強(qiáng)度和耐電壓的測(cè)定方法》
4、ASTM D149《固體電絕緣材料工業(yè)電源頻率下的介電擊穿電壓和介電強(qiáng)度的試驗(yàn)方法》
三、功能:
1、試驗(yàn)過(guò)程中可動(dòng)態(tài)繪制出試驗(yàn)曲線,試驗(yàn)的曲線可以多種顏色疊加對(duì)比。
2、可對(duì)試驗(yàn)數(shù)據(jù)進(jìn)行編輯修改,靈活適用;
3、試驗(yàn)條件及測(cè)試結(jié)果等數(shù)據(jù)可自動(dòng)存儲(chǔ);
4、試驗(yàn)報(bào)告格式靈活可變,適用于不同用戶的不同需求;
5、可對(duì)一組試驗(yàn)中曲線數(shù)據(jù)的有效與否進(jìn)行人為選定;
6、試驗(yàn)結(jié)果數(shù)據(jù)可導(dǎo)入EXECL,WORD文檔編輯;
7、過(guò)電流保護(hù)裝置有足夠的靈敏度,能夠保證試樣擊穿時(shí)在0.1S內(nèi)切斷電源;
8、儀器運(yùn)行的持久性: 儀器可連續(xù)運(yùn)行使用,不需為保護(hù)儀器而定期停機(jī)。
四、主要技術(shù)參數(shù):
1、輸入電壓: AC 220V
2、輸出電壓AC: 0~50KV;DC:0~50KV
3、輸出功率: 5KVA
4、測(cè)量范圍: 0KV~50KV
5、電壓測(cè)量誤差:≤ 2%
6、升壓速率: 0.1kV/s ~10 kV/s
7、耐壓試驗(yàn)電壓:10~50KV連續(xù)可調(diào)
8、耐壓時(shí)間: 0~4H(無(wú)電流導(dǎo)通情況下)
9、電源: 220V±10%的單相交流電壓和50Hz±1%的頻率
10、電源電壓穩(wěn)定度外界電源電壓波動(dòng)≤10%
11、長(zhǎng)×寬×高700mm×800mm×1650mm
12、設(shè)備自重: 100kg
13、運(yùn)行環(huán)境溫度:15 ~ 30℃,相對(duì)濕度:30%~85%能夠穩(wěn)定運(yùn)行。
14、接地要求儀器需要單獨(dú)接地,接地附合國(guó)家標(biāo)準(zhǔn)要求
15、接地電阻要求≤4Ω(用戶實(shí)驗(yàn)室自行準(zhǔn)備)
四、主要配置:
1、主機(jī) 一臺(tái)
2、試驗(yàn)臺(tái) 一個(gè)
3、油盒 一個(gè)
4、試驗(yàn)電極 三個(gè)
5、自動(dòng)放電裝置 一套
6、試驗(yàn)軟件 一套
7、計(jì)算機(jī) 一臺(tái)
8、噴墨打印機(jī) 一臺(tái)
9、電源線數(shù)據(jù)線 一套
五、絕緣材料介電擊穿強(qiáng)度測(cè)試儀安全說(shuō)明:
1、要安裝單獨(dú)的保護(hù)地線,主要是減少試樣擊穿時(shí)對(duì)周?chē)a(chǎn)生的較強(qiáng)的電磁干擾。也可避免控制計(jì)算機(jī)失控。
2、直流試驗(yàn)放電保護(hù)功能:在設(shè)備做完直流試驗(yàn)時(shí),當(dāng)開(kāi)啟試驗(yàn)門(mén)時(shí)設(shè)備會(huì)自動(dòng)保護(hù),直至使用設(shè)備上的放電裝置放電后保護(hù)會(huì)自動(dòng)取消.(注:因?yàn)橹绷髟囼?yàn)后不放電會(huì)危險(xiǎn)到人身安全,不能直接拿取電極,起到提醒使用人員放電以免造成傷害)。
3、增配試驗(yàn)手動(dòng)放電裝置,隨主機(jī)為一體化,當(dāng)直流試驗(yàn)過(guò)程中突然斷電,可采用手動(dòng)放電棒進(jìn)行放電,保證試驗(yàn)人員的人身安全。
4、該試驗(yàn)設(shè)備的電路設(shè)有多項(xiàng)保護(hù)措施,主要有:過(guò)流保護(hù)、失壓保護(hù)、漏電保護(hù)、短路保護(hù)、直流試驗(yàn)放電保護(hù)等。
5、六級(jí)高壓安全斷電控制:
1、總電源開(kāi)關(guān)
2、高壓斷電開(kāi)關(guān)(鑰匙開(kāi)關(guān))
3、調(diào)壓器復(fù)位開(kāi)關(guān)
4、試驗(yàn)箱門(mén)安全開(kāi)關(guān)
5、高壓變壓器輸入端限流空開(kāi)
6、漏電保護(hù)開(kāi)關(guān)
英文介紹:
First, the use of
ZJC-50KV by computer control of medium frequency voltage breakdown of insulation, power frequency withstand voltage test is completed through man-machine dialogue. Mainly applies to materials such as insulation paint, resin and rubber, impregnated fiber products, mica and its products, plastics, composite film products, ceramics and glass medium in the frequency of the voltage or DC voltage breakdown strength and voltage time test resistance in Solid Electrical; controlled by computer, all the test data of rapid and accurate acquisition, processing, and access, display, print.
Two, according to the standard
1, GB1408.1-2006 "electrical strength of insulating materials test method"
2, "GB1408.2-2006 electrical strength of insulating materials test methods
Part second: Additional requirements for tests using direct voltage "
3, GB/T1695-2005 "vulcanized rubber power frequency breakdown voltage strength and resistance to voltage measurement method"
4. D149 ASTM test method for dielectric breakdown voltage and dielectric strength of solid electrical insulating material industrial power frequency
Three, function:
1, the test process can be dynamically drawn out of the test curve, the curve can be compared to a variety of color overlay.
2, the test data can be edited and modified, flexible;
3, test conditions and test results and other data can be stored automatically;
4, test report format flexible, applicable to different needs of different users;
5, can be used for a group of test data in the curve of the effective or not artificially selected;
6, the results of the test data can be imported into WORD, EXECL document editing;
7, over current protection device with sufficient sensitivity, to ensure that the sample breakdown in the 0.1S cut off the power supply;
8, the persistence of the operation of the instrument: the instrument can be used continuously, without the need for the protection of equipment and regular downtime.
Four, the main technical parameters:
1, input voltage: AC 220V
2, the output voltage of AC: 0 ~ 50KV; DC:0 ~ 50KV
3, output power: 5KVA
4, measurement range: 0KV ~ 50KV
5, voltage measurement error is less than 2%.
6, pressure rate: 0.1kV/s ~ 10 kV/s
7, voltage withstand test voltage: 10 ~ 50KV continuously adjustable
8, pressure time: 0 ~ 4H (no current situation)
9 power: 220V + 10% 50Hz + 1% single-phase AC voltage and frequency
10, the power supply voltage stability of the external power supply voltage fluctuation is less than or equal to 10%
11, length * width * high 700mm * 800mm * 1650mm
12, equipment weight: 100kg
13, operating environment temperature: 15 ~ 30 degrees, relative humidity: 30% ~ 85% can be stable operation.
14, grounding requirements instruments need to be grounded alone, and then meet the requirements of the national standard
15, grounding resistance less than 4 ohm (user laboratory preparation)
Four, the main configuration:
1, the host a
2 test a
3, oil box a
4, the test electrode three
5, automatic discharge device
6, a set of test software
7, a computer
8, a inkjet printer
9, a set of power line data
Five, safety instructions:
1, to install a separate protective ground, mainly to reduce the breakdown of the sample around the strong electromagnetic interference generated. Can also avoid the control of computer control.
2, DC test discharge protection function: when we finish DC test equipment, when opening the door test equipment will automatic protection until use equipment discharge device for discharging protection will be automatically cancelled. (Note: because the DC test on it would be dangerous to personal safety, can not be directly fetching electrode to remind users discharge so as not to cause harm).
3, equipped with test manual discharging device, with the host for the integration, when DC test process in case of sudden power failure, can use manual discharging rod discharge, ensure the safety of testing personnel.
4, the circuit of the test equipment is equipped with a number of protective measures, mainly include: over current protection, voltage loss protection, leakage protection, short-circuit protection, DC test discharge protection, etc..
5, six level high voltage safety power control:
1, the total power switch
2, high pressure switch (key switch)
3, voltage regulator reset switch
4 test box, door safety switch
5, high voltage transformer input limit open flow
6, leakage protection switch